Today I'll be giving a talk on mechanical ventilation on obstructive legacies. I do have a PDF of this talk, and this is the QR code. I think it should work for everybody to actually view and view for yourselves if you want. If any issues with that, there's my email and also my number. I have to actually send it or re-upload it. Okay, so most of the talk today is going to be on invasive mechanical ventilation and the approach to waveform analysis.
and respiratory mechanics in general, with special emphasis on auto-peak. I will touch very, very lightly, you know, non-invasive ventilation is particularly important in respect to one disease, particularly COPD. However, I will also be probably focusing more on the COPD asthma concept of obstructive lung disease. There are other disease processes that cause obstruction, of course, but most of the literature, as we know, is dedicated to both of these disease processes. So I'll start with a case. This was a case from maybe two to three weeks ago that came to our ICU at the hospital center. This is a 35-year-old man, a history of asthma, and he is a smoker. He's brought in by the ambulance because he was complaining of the two torts of breath. He was tripod-ing an extramus. He had a silent chest. He was bitten several.
medications in order to try and help him out. When he came to the ED, his temperature was okay. He was having an abnormal vital sign. His heart rate was low on the lower side. Obviously, the rest of the weight of stress, his breast cancer rate was in the 40s. Blood pressure, 220 over 90, it's not okay. But as soon as he got there, he lost consciousness and was enlarging and intubated. Bhavik, I'm just coming back to the QR code for you. I also will have it at the end of the talk as well for those who missed it. So you don't have to worry.
Hope this helps. Okay, so back to our patient. He gets a blood fraud. Things are not particularly abnormal. He has an EKG that shows that he has COVID-19 and his urine drug screen is positive for THC. And by history, actually, his partner actually disclosed that he was actually smoking marijuana before this all happened.
So, in the ED, these are his initial ventilator settings. He's put on a respiratory rate of 6, tidal volume of 450, baseline of 100 percent, just as the initial setting, and a peak of 5. This peak at the time was 26, plateau of 15. There's no auto-peep, and then a post-intubation ABG is actually collected. It shows severe respiratory acididosis, pH of 708, TCO2, and oxygen level of 296.
So the talk today would be how to approach a patient that comes into your ICU looking like this. I'll talk briefly about the trend in mechanical ventilation in obstructive lung disease actually over the past two decades. This is a cohort of 24,000 patients from 38 countries. You'll notice that there's been a decreased rate of acute exacerbations of COPD among mechanically ventilated patients. So it's down from 10% in 1998 to about 8% in 2016. Since then, there's been an increased use in non-invasive ventilation as the first ventilatory support in the ICU, and a decreased overall mortality, as you can see in the bottom.
But when we talk about COPD in particular, there's been a more recent study that looked at what the epidemiology truly is. And it looks like most of the patients that do have a diagnosis of CP or COPD actually get intimidated for other causes rather than acute exacerbations. Only 12% have been found to have acute exacerbations. And when we looked at these patients, it was actually favorable over all outputs. They had shorter mechanical ventilator days.
like I was saying, and their mortality from the ICU in the hospital was also better. So it looks like ventilating acute exacerbations for COPD invasively is overall big. But before we actually talk about the actual ventilatory strategies, I'm going to spend some time talking about respiratory system mechanics in obstructive lung disease and also touching highly upon the concept of model P-pathal dynamic hyperinflation because these are the two pathologies that are probably the most important when it comes to obstructive lung disease in general.
So if you guys remember, positive pressure ventilation, it is dependent on Bohm's law. We need two pressures at two different points, and the difference between these two pressures will allow gas to flow and eventually distend the loss. So here you have the equation of motion, which dictates that the pressure that the ventilator has to give during installation has to overcome resistive forces of the airways in the tubing, which is your resistive pressure, and then reach the lung. distensive which is the distensive pressure and there's always usually a baseline of PEEP and your alveolar pressure is of course equal the pressure that has been used to actually generate the volume plus the PEEP that the alveolar provides. Now isolation is a little bit different because here it is completely Technically, the ventilator shuts off, and now the flow of gas from the lungs back to the atmosphere or to the ventilator is dependent on the alveolar pressure and its difference between that and the airway pressures. So you'll notice that the distensive pressures and the resistive pressures actually equalize in order to generate that flow. Another important concept when it comes to exhalation is natural decay, or the time constant.
So the lungs actually follow the natural processes of losing air, and there is a specific time by which a certain amount of lung volume or pressure is lost from the lungs. So this, of course, assumes that the lung compartment model would be a single resistance. I personally prefer memorizing the numbers for each time constant. You'll notice 37% of the volumes left in the 1st after 1 time constant is 40% by 2. And you'll notice that 4, 2% of the long volume.
This is just a big figure of what normal passive ventilation would look like. This is different modes, waveform analysis based on different ones. Of course, it's different based on a lot of mechanics with the ventilator settings are, the patient characteristics, and of course, the interaction between the ventilator and the patient himself. Now, when we look at normal passive inhalation and exhalation, I can hear the different phases of the respiratory cycles. You start inspiration at point A. The pressures of the ventilator are high. The pressures of the alveoli equal the PEEP. This difference in pressure generates flow. And in inspiration, the pressures of the alveoli and the airways actually equalize. And then at start expiration, now the ventilator cycles are worse. So it goes from a positive pressure that was used to deliver the volume down to a PEEP. And the difference between the alveolar pressures and that PEEP will generate the flow.
And towards the end of expiration, all of the lung volume is exhaled once the pressure is equalized and we return back to our resting volumes, or the FLC. Now, we've got increased resistance. Again, depending on the load, things will be different when we look at the waveforms. When we talk about a volume control, we talk about the increase in the P pressures because the resistive pressures went up. But almost universally, during exhalation, you'll notice that there is decreased load because of that increased resistance and an increase in the time constant, or the RC.
regardless what the mode of ventilation is used. In pressure-controlled ventilation, you'll notice that the resistance goes up. The result of decreased flow also leads to a reduction of the overall inhaled volume. But again, on exhalation, decreased flow increases. Here's a picture of waveforms in front of the ventilator, and you'll notice how the key pressures are exceeding the time here, and I'm saying that key bronchospatine and increased respiratory resistance. And this figure also depicts better the increased time constant, so you'll notice that it takes longer for the lungs to actually return to resting volume, and depending on the disease process, you will have to get a longer oxygen.
Sometimes when you look at the better layer itself, you'll notice that the RC is particularly long, just by subjective observations. You'll notice on the bottom side curve here at the bottom, the return to baseline takes a very long time and the curve is kind of pulled upwards from the middle. Now, the most important pathology, like I said earlier, in obstructed lung disease of severe obstruction, in particular, is dynamic hyperinflation and auto-PEAP. Now, dynamic hyperinflation is a volume phenomenon where there's not enough time to actually exhale, so there is residual lung volume at the end of exploration. What results from that is a pressure phenomenon, or auto-PEAP, also called the intensive PEAP, where the end expiratory alveolar pressures are higher because of that routine.
What basically happens here, again, I'm a visual person, so I like to actually look at things visually. You'll notice at the end of inspiration, the alveolar and templatory pressures have equalized. We've got the ventilator cycles, but the problem is now we have increased resistance. So the flow rate is lower, and there is decreased volume that leads along by the end of exhalation. So because of that retained volume, now you have a certain degree of bottom peak. That's the number five in red.
in the lungs. And when ventilation starts again, you'll notice that the machine now has to apply extra pressure in order to overcome not only the resistance, but also overcome the auto-PEEP that is in the lungs to deliver the same tidal volume. So at the end of the next breath, you'll notice that the alveolar pressure is much higher than where they started in the first one. Again, this is a depiction of several breaths. with increased resistance and the formation of dynamic hyperinflation. So you'll notice with every start of exhalation, the difference in pressure, the trans-airly pressure, the driving pressure, if you may, between the ventilator and the alveolite continues to increase in such a way that it eventually reaches that equilibrium. So you'll notice that this is why nobody truly...
pops their lungs if they continue to stack air. The air stacking has to stop at some point, and that's where the equilibrium process is reached. And the reason for that is now that the delta P between the lungs and the alveolite at the start of exhalation becomes so high that even though the time that is required to exhale is low, the volume that is generated by this large delta P becomes equal to the tidal volume that was actually set. So we have an equilibrium of both auto P and dynamic hyperinflation as well. So what is the problem that happened with auto-peakment and hyperinflation? And controlled breaths in general, you worry about the high pressures that the ventilator actually has to produce. So your peak pressures are high, and if the peak volume is too high, the plateau pressures may go up as well. And we all know that the plateau pressures and the alveolar pressures are more important in resulting in bad outcomes. The most important complication that we tend to aim to avoid, of course, is aerotrauma or human genetic complications.
Now, not all of our patients are completely passive. Some of them are going to be working as well. And the presence of auto PEEP is known to increase the work of breathing, particularly with patients trying to overcome this intrinsic PEEP. A lot of times you will see an effective triggering, which we'll touch on a little bit later. The larger the lungs, the worse the length of tension relationships, so it's more difficult to take a breath when the lungs are fuller. So if you try it yourself, take a very, very deep breath, and then try to take a breath on top of that, it's going to be much, much harder. And of course, respiratory muscles because of the high pressure, so you have less percussion, and it causes muscle damage as well.
We've all seen pressure volume curves of the lung. This is also a depiction of what you may see if there is intrinsic PEEP. So you'll notice in the green is your normal tidal volume ventilation. The patient needs about a negative two centimeters of water to trigger the vent, so not a lot of work that is required here. However, when patients who have auto PEEP, you'll notice now that your resting lung volume rises dramatically. The pressures of the lung also rise dramatically.
And now in order for the patient to trigger, they have to generate a very, very large negative inspiratory force in order to reach that trigger point and generate the volume. And you'll notice also since the pressures of the volumes are higher, we're moving upwards in our PD curve. And even though the tidal volumes have not changed, we will very easily surpass our upper inflection point because of the microscopic and necroscopic barotone. This is some data from about 10 patients that were compared, so 10 COPD patients that were compared to 18 healthy patients. So look at what the work of breathing that was generated by each set of patients was, and you'll notice that...
Despite the fact that they were all paralyzed and they had similar ventilator settings, there was a lot of resistant work that was generated by those patients from COPD just because of the obstruction, and also work that was needed to overcome the intrinsic piece itself. So a dramatic rise in the work of bleeding when you have obstructive lung disease in general. Okay, so let's talk a little bit more in depth about auto-PEAP. The landmark paper that actually first described it was in 1982 by Pepe Molini. They saw it in three patients, two had COPD, and one had thermal injury prior to coming to the ICU. Autopeep is seen in every single mechanically ventilated patient with COPD. It's also been reported in about one-third of the patients without COPD history. The causes are outlined in front of you. The two big ones that I'll be talking about are decreasing expiratory time, but probably the most important cause of autopeep when it comes to obstructive lung disease is an increase in their expiratory time constant because of that increase in resistance.
Other causes, if people are actually trying to inhale during exhalation, they may not exhale the full amount of volume that they're supposed to, so sometimes they will develop auto-peak despite the lack of increased resistance or obstruction. Also, a high tidal volume will result in more residual air in the lungs based on your natural decay curve. And also, any external resistance to the flow, such as MET2 or extreme patients, for example, may have problems as well. So, normally... In healthy individuals, the normal time constant is about 0.2 seconds. So if you give somebody about one second to exhale, they should be doing okay, because they need four more time constants to complete the exhale. But in clinical practice, we generally say that we see three time constants, which are acceptable in one solution.
Great. There's a lot of causes for decreasing the expiratory time. Some are patient-generated because of their breathing pattern, but be aware that value of settings in particular can also cause that. So an increase in the frequency of the respiratory rate decreases that exhalation time, and also increase in the inspiratory time may result in that decrease in the exhalation time as well. So any end expiratory pauses will increase your inspiratory time, and also low inspiratory flow rates will increase the eye time for any given volume.
Now, when we talk about increasing the time constant as a cause for auto-peep, I always have to mention that since auto-assets of time constant is a result of resistance, not compliance, both an increase in resistance or an increase in compliance can cause auto-peep. Although less common, keep in mind to some of those patients, for example, very, very nice, compliant ones because it's just one big boolean, basically, of the destruction of the album. And because of this high compliance, the driving pressure, it results in excellent response. the gas actually goes down. So you'll notice in the figure, the plateau pressures are about 15 compared to 25 or something. So that leads to the pressure that is lower in COPD, which results in a slower exhalation time. It sometimes can develop auto-PEP or be something that adds to the auto-PEP and COPD conditions. But this is probably the most important cause in the patient population that we're talking about. So increasing the resistance because of the abnormality in the air.
There are two big types that I'll be talking about. The omic resistance is a simple one. It's a fixed decrease of the airway diameter. Our asthma patients, you'll see that in our asthma patients, obstructive bronchiolitis and any upper airway obstruction that's fixed. Keep in mind asthma, the increased resistance is asthma because of airway inflammation and bronchospasm. Whereas in COPD patients, there's the concept of expiratory flow limitation, which is airway narrowing. It happens during exclamation because of airway collapse, as you can see in the figure.
I'll talk a little bit more about this during flow of limitation. So it always occurs during forcible exhalation, even with healthy subjects. But it can occur during restful breeding as well. So when that happens, there's usually pathology that's involved with the small airways that makes them tend to cave in or collapse. In COPD, there's loss of the elastic scalability surrounding these airwaves, but you may see dynamic airway collapse also in other topologies, such as ARDS because the lung rates are high, or you may see it also in OVC because the chest wall weight is also high.
Some people have looked into this a little bit more closely and noticed that there may be some concepts of wave speed limitation, wherein the airways themselves, when they are subjected to high pressures that make them collapse, they can't really adjust the airflow and the velocity of the air that's moving as the pressure travels along those airways. And this, of course, all results in wind. a true air trapping. So in COPD, we see what is termed as true air trapping, and that's because the expiratory to inspiratory resistance actually, that's a significant difference. So you'll see the ratio between both becomes now higher. And there's been some studies that support that if you measure this respiratory to inspiratory resistance, it can predict the true presence of obstructive lung disease.
Now, expiratory flow limitation, there's several ways that you can actually detect it on the ventilator itself. Probably the simplest one is if you look at the expiratory flow curve, then you'll notice that there's an initial maximal increase in that expiratory flow, followed by a dramatic decrease, so very similar to your smooth appearance on the flow volume curve in CO2 patients. But also, there's other ways that you can actually try and detect expiratory flow limitation, and that's with the application of external people. And that is due to the waterfall effect, which I'll talk about in a couple of slides. So notice here, very, very dramatic initial rise in the flow during exhalation, followed by severe reduction in that flow. So that just tells you that the airways are open, and then the pressures became high enough for the airways to collapse, and the flow rate goes down, that's what the flow rate goes down. Now the waterfall effect has always been something that happens.
I found it a little bit difficult to try and understand. I'll try to make it as simple as possible. The autopie that is within the lungs is transmitted to the rest of the pulmonary system that's interthoracic. So basically, if you look at this example, autopie is 15 on the left side. So the external pressure upon the airways is also 15.
Any application of PEEP that is below this number will not cause any change in the lungs and will not be transmitted into the lungs at all. So you'll notice as we go up on the PEEP and finally reach a PEEP of 20, which exceeds that critical pressure that's blocking those airways, now those airways open up and that PEEP is translated to the lungs and can actually worsen the total PEEP that we have. And here's a very nice representation of the waveforms comparing expiratory flow limitations at the top versus auto-PEEP that is not caused by expiratory flow limitation, which is probably more related to olig resistance. So you'll notice as you go up on the PEEP, the peaks don't change unless you bypass or supersede that auto-PEEP level of about 10, right? As compared to the graph in the bottom, you'll notice in olig resistance patients as you go up on the PEEP, both the plateau and the PEEP pressures will go up.
So can we predict auto-peep by looking at the waveforms? And the answer is yes, and you guys probably know all of these signs in general. So a mismatched area under the inspiratory and excitatory flow curve is on the other side of this. Keep in mind that the area under the flow curve is the volume. So if the volume during acceleration as depicted by the area under the curve is smaller, then this may signify auto-peep. This is a more dramatic example of this, so you'll notice how small the area under the curve on acceleration is here compared to the inspiratory area under the curve.
You see this also in the second and third breath in this example as well. And you can safely say that something is odd peeping if you see it this dramatically. The other side that you may see is persistent and expiratory flow, right? So the presence of flow at the end of exhalation means it belongs to that MTB. So you see it clearly here in this example and also another example with pressure control of persistent expiratory flow before expiration starts.
One thing that's important is to recognize that the degree of persistent flow does not predict the degree of auto-peep, because keep in mind the ventilator settings are different for each and every individual. So there is no correlation between that degree of and how bad the auto-peep can be. And then the third sign that may predict that auto-peep is present is... So an effective trigger dictates that the patient generates some force to try to trigger the ventilator. Universally, we are taught that auto-treat causes this. But keep in mind also, people who are weaker, who've been ventilated for a really long time, may actually have very weak diaphragms to try and trigger the ventilator. The way that you see it is generally a reversal in the flow with no...
resultant triggering of the vent. There's no breath that happens. So you see it very nicely here. The flow actually reaches zero. The patient actually tried to take a breath, but was not rewarded with that breath. In this case, you'll see that the patient actually is generating a little bit of effort that is longer than the previous example. You see two subsequent breaths with no reward. This one is a little bit less dramatic.
in terms of phosphoridation. But this one is very clear. You'll see that the patient actually reversed the flow very nicely and even was able to change the positive flexion and the pressure reform as well. Now, can we measure our peak? And the answer, of course, is yes. You guys know that an expiratory positive invert will lead to equilibration of the alveolar pressures and the airway pressures at the end of exhalation. And if there's any auto-PEEP, that number should be higher than your set PEEP for the ventilator. And the difference between that total PEEP and your set PEEP will be the value of auto-PEEP itself.
The screen actually shows you the pause itself. So you'll notice here the total PEEP is about 12. The set PEEP is about 5. And you'll notice that the intrinsic PEEP is around 7. Now keep in mind, this maneuver can be a little bit inaccurate if the patient is generating any effort during the maneuver. In the past,
When, for the older ventilators, when somebody would actually generate this maneuver, there would be gas compression with the valve fusion, and somebody would sometimes overestimate these measurements. Our more moderate ventilators, of course, now avoid that because we have the pneumatic valves at the airway opening. Similarly, we can actually measure the volume that's resulting in dynamic hyperinflation if you do an aspiratory depositator. However, the biggest difference here is we're actually allowing the patient to completely exhale, so we're opening the aspiratory valve here. This is the concept that you see when you actually disconnect to the patient from the circuit when they're obviously so bad and causing them to have a compromise.
So you allow the amount of air to actually escape. And you'll notice that all of that dynamic hyperinflation volume now leaves the lungs. This is another example of tracings in somebody on a square wave form. The respiratory rate is actually decreased dramatically down to one breath per minute. Not a lot of ventilators would actually like that. So you'll notice that the end expiratory volume now goes down by 0.8 liters, which is the amount of air trapped.
but again notice once the original respiratory rate is returned dynamic hyperinflation again occurs after about nine breaths and the total one volume is 1.2 meters which represents the dynamic hyperinflation volume plus the timing volume. There's other ways to actually measure auto-peeps. So an expiratory positive over is considered a static method, but there are dynamic methods where you can try and predict what the intrinsic peak is. So if you look very, very closely at ventilator waveforms, and notice how much rise in the airway pressure there is before the generation of flow, that usually is representative of the amount of pressure that the ventilator had to produce in order to overcome that intrinsic peak. So that is one method, if you're able to see it, that's always useful.
However, in patients who are actually breathing, it's a little bit difficult to get your static machines, right? So remember, the ventilator is a computer. If you're going to get an respiratory pause button, it's actually going to wait until it ends the respiratory cycle. So if you say, if somebody had a respiratory rate of 12, then the machine is going to count up to five seconds, which is the length of the respiratory cycle, before it actually applies the pause itself. Now, if anybody's breathing, at a rate that's higher than the vent, then usually the respiratory pause maneuver cannot be personalized. And that's why, in a lot of cases, you'll see if there's a cancellation of the user because of the patient is actually breathing.
So the best way to try and actually calculate or measure the odd treatment of these patients is usually with the use of esophageal blood. You actually measure the pleural pressures in the blood and see how much there is of a drop in pleural pressures or esophageal pressures before the initiation is low. And that degree of drop, again, is representative of the force that the patient has to generate to overcome that intransigrate. Now, calculating the time constant can also be very, very, very important when we're dealing with a severe restructuring for these patients because the respiratory rate can be determined, or the best respiratory rate can be determined for these patients to try to minimize that intensive pre-calculation error. There are several cases that you can actually do this. Probably the most accurate way is to measure both the resistance and the compliance of the patient, and this is best done, of course, with a constant flow volume controller where you can look at, where you can actually measure your plateau pressures, measure your peak pressures, and based on your equation of motion, you can come up with both your compliance and resistance values.
and then your tau or your time constant is going to do the products of both of these two values. In this example, get a compliance of about 0.05 and resistance of 16 and generates a time constant of 0.8 seconds. Using volume control decelerating rule is not the way to actually get these numbers. Let's keep in mind that because the flow is decelerating, the resistance will be difficult to measure in this case. The other important method of doing this is to actually go by a perfect way. So your natural decay curve, we know that
you have only 37% of the remaining volume after one time constant. So if you actually measure from your tidal volume down to 37% of that volume and see how much time there is, you may be able to at least estimate the time constant. So this example is about 2.4 seconds. And this is actually back to our patient that I presented at the very beginning. So he's on tidal volume 700, which is your 100%. And I jumped out the time where we actually get that number.
And then you calculate basically 37% of the 700 TCs. And you look at what the time is at that point in particular. So it's about 1.4 seconds. And generally I tell people try to do it at multiple places. Try to get your second time constant as well. But notice in this case that there is somewhat of a discrepancy in the measured time constants. And that's because this patient is not completely as passive as you want them to be. And as I'll explain in a second, there's a lot of heterogeneity that prevents the numbers to get as pristine as you want them to be.
A third method is actually also looking at your equation of motion. We're looking at your Ohm's law equation during exclamation. So if you guys remember, the size of pressures will equal your resistive pressures. So your F times R equals your volume of compliance. So if you reorient the equations, you get that the tau, which is the resistance times compliance, is equal to the volume of over flow. And then the negative sign right there just denotes that the flow is happening in the opposite direction than inspiratory flow.
and I generally say try to get as many points as you can. So here's an example for our patient. And you look at what the volume is and what the corresponding flow is. But be very, very careful when we're actually doing this because you want to make sure that your minutes are correct. So in looking at long mechanics, we use liters for our volumes. And when we look at the time, we're always looking at seconds and not at minutes. So you have to make sure that the emergence and you will get an approximation of what the appropriate time function is. So once you get these numbers, this will give you a good understanding of what your expiratory function is. And like I said, three or more taus will generally allow for an acceptable exclamation.
Now all these maneuvers. Whether it's measuring the intrinsic PEEP, whether it's measuring the dynamic of hyper-insulation, or the time constant, they've all assumed that the Lung functions as a single compartment. However, this is never the case. There's a lot of inequalities that have to change in the Lung. So different Lung units will have different time constants, they may have different intrinsic PEEPs, and different maintained volume. So be very, very careful when you're looking at these when you're doing these with your first step. Make sure you understand that things are probably different than the Lung. You might see them on your calculations or measurements. Dynamic measurements of Auto-PEEP in particular, like when you're looking at the rise of pressure or the drop in the esophageal pressure before flow starts, usually underestimate the magnitude of Auto-PEEP.
And the reason for that is because sometimes all you need is a small amount of pressure to allow flow in the best one units. So now that we looked at maybe 10 patients, notice that the static PEEP in these patients was sometimes much higher, even as much as 9% higher than your dynamic PEEP. So that's why generally use the dynamic PEEP includes a lot of caution. Okay, so really quick for the absence of non-invasive ventilation. It took about a decade for people to actually accept the fact that people actually do well in COP. It was actually first described in the 60s and it wasn't until the 90s that we were able to generate studies to improve mortality and decrease inhibition of NIP in general.
We know for a fact what I said, it decreases mortality, decreases intubation rate, but there's also been some data supporting that if you fail your non-invasive ventilation, then the mortality goes up. However, there is a little bit of asterisk in that, because they actually just know that somebody's really, really sick and that their mortality was never really different than those who were first starting on invasive mechanical dentistry. There are some clinicians, particularly in Europe, that actually propose that YPAP or NIV B, the ceiling, are the most that you actually provide patients in terms of inventory assistance. The data comes actually from a lot of do not intubate patients, but they see very, very good outcomes in these patients who are not intubated. When it comes to asthma, though, things are a little bit different. Most of the studies that we have a little bit perspective, they may be supportive of NIV, however, there's no high quality of trials to support this, and currently the guidelines still recommend not in case of ventilation in asthma. So, let's talk a little bit about...
what we do for our patients. When should be considered? When should we intubate? The answer is you guys are all a little bit experienced now in trying to understand when something needs to be intubated. So this is the only indication for intubation is thinking about it. The true, absolute indication for intubation is cardiac arrest and respiratory arrest. Of course, if you have somebody on BIPAP or non-invasive evaluation, after two hours, things are not looking good, particularly in your pages, while your consciousness is getting worse, or during this between stress, these are the patients that should get intubated in the channel.
And does respiratory support actually improve things? The answer is yes. Again, whether it's not invasive or invasive, you can decrease respiratory effort and decrease the work of breathing patients. These are 11 COPD patients that were studied with two different modes, so intermittent ventilation and pressure support ventilation. And you'll notice the more we actually provide support, regardless of what the ventilatory mode is, the better your work of breathing, which is reflected by the pressure time products in the left graph, and also improvement in the dyspnea overall.
What mode should you use? The short answer is it doesn't matter. The long answer is know your ventilator modes, understand how they work. Whatever you're comfortable with generally should work. Most authors actually support that you use assistive tool, volume or pressure. It's better than pressure support ventilation because sometimes you need to sedate the patients you don't care less. Some authors say to use constant flow because you can monitor low mechanics a little bit better.
There's some studies on PRDC comparing them to SIMD. PRDC may have been a little bit better, but then again, these are small studies. There's a lot of talk about AVA and proportional assist ventilation as well to help in the patients, but there's no good data to support using them for all patients. Try to avoid control of those because you will, if you're basically a patient, and may increase the work for it. And APRD is an absolute contraindication in obstructive lung disease. That's in the worst because a lot of people are already there and worst internet.
Located volume versus pressure. Again, most of the reviews and system meta-analyses have shown that there is no real difference between those, whether it's compliance, gas exchanging, dynamics, or work of reading, but also specifically no difference in ICU. What ventilation pulse should be targeted? Whatever targets you have for every single ventilator should still apply here. So your traditional goals of ventilation and oxygenation apply. So make sure that these are met. However, if you want to go a little bit more in depth in terms of obstructive body disease, the four major goals are decreased respiratory effort, decreased work of breathing, minimize the auto-peak, make sure that respiratory acid is really bad, and of course, maintain oxygenation.
Now these four goals, prevention of hyperinflation and auto-peak is the complications that come from this slide particularly here. So we want to prevent circulatory collapse. And of course, from the previous slides that we actually discussed, there are several assessments that we can do. So look for the science, try to quantify, try to assess what the system is and try to calculate the time cost if you can. I'll pass for a question. I may run maybe five minutes over time, but you have a 45-year-old woman with a UPD classification. She's on pressure support.
This is her graphics. Just based on these waveforms, what change would the patient most benefit from? I have a chat box so you guys can type it in the chat box if you would like. I'm also happy for you guys to shut it up. Okay, so if your answer is PBs and Cs, I will say yes or you are correct in this one. So notice that the PEEP is zero here.
And there are signs of ineffective triggers that you see right there. And the question is, would applications PEEP actually be helpful? A small amount of PEEP is actually useful in our obstructive lung disease patients because it actually stores ventilation to the dependent regions of the lung and may prevent the collapse of the airways in general. So that's why we try not to try to avoid a PEEP of zero, although we're hoping that this would actually generate a driving pressure. So the other big and important concept is trigger.
Right, so the higher your PEEP, the less the efforts that the patient has to generate in order to trigger the ventilator in case of ineffective treatment. So you'll notice in the diagram or illustration here, the patient generates a good amount of effort, drops the alveolar volume high enough for them to drop the airway pressures enough to trigger the ventilator. bring up that peep can decrease that delta that the patient actually has to generate and will result in effective triggering despite the fact that the auto-peep itself never actually changed. So we generally say because of the waterfall effect.
Don't give external PEEP that is more than the intrinsic PEEP because then you will worsen total PEEP altogether. Most authors support 70% to 80% of the intrinsic PEEP. Apply externally as appropriate. But again, remember, external PEEP will not treat dynamic hyperinflation and will not improve the actual intrinsic PEEP. Another question.
My five-year-old guy with asthma is intubated in the ICU. Blood gas shows respiratory acidosis. He's given a paralytic. He's placed on a volume control ventilation. Respiratory rate is 30. Thinal volume is 400. He's 5. And 5 and 2 is 100%. The respiratory therapist is censored. And the nurse is noticing a drop of blood pressure. Which is the best intervention here? To increase sedation, give more paralytic, increase the respiratory rate, decrease the respiratory rate, or switch to pressure control. All right.
We see a lot of Ds, and D is the answer. So is decreasing the respiratory rate beneficial? The answer is yes. So I have here different people with different respiratory rates, and then we change the respiratory rates by decreasing it. And you'll notice going from 15 to 6 has a dramatic improvement in the respiratory rate altogether. And there's very, very good data from patients supporting that if you drop the respiratory rates, you actually improve some mechanics. The pace might not change.
because of the resistance itself. But notice, both your plateau and your auto-peep levels will go down if you go down lower. And also, the end-experts will also be better if your respiratory rates are lower. This is a, again, waveform analysis in somebody with a higher respiratory rate and converted to a lower respiratory rate. Notice the amount of auto-peep goes down from 5.5 to 1.9. And also, the plateaus improves because the incidence of the peep is lower, so it goes down from 10 to 6.2.
In our patient that I actually, in this case I presented initially, this is different representations of different respiratory rates. So you'll notice that the left side, the rate is 22, with intrinsic people about 4. Throughout the rate to 18, the auto-peak gets better to about 2. But if you drop the rate to 10, there's almost no auto-peak whatsoever. What about decreasing the respiratory time? So it is not as beneficial as decreasing the respiratory rate overall in order to decrease your respiratory time. There has been a lot of studies looking at is there
any detrimental effects of a higher inspiratory time. In these waveforms, you'll notice when patients are trying to take a breath, sometimes a longer inspiratory time that's higher than their neuropi time may lead to what we call a trigger excess, which could lead to more increased work of breathing and cycle distinctiveness. So this is looking at 10 patients with COPD on pressure support ventilation. They have different expiratory trigger settings that actually decreased their expiratory time. So 10%, 25%, 50%, and 70%. And you'll notice that there is better ventilator synchrony decreased ineffective triggers. So this was an ineffective trigger, ineffective trigger. This is an ineffective trigger. They go away once you decrease the expiratory time.
However, there's also big subsequent data that show that if you decrease the respiratory time, the patients automatically start increasing their respiratory rate. And this is independent of the flow or the tidal volume. And it's thought to be because of something called the Herring Brewer reflex, where the neural eye time going down eventually also leads to a decrease in neural e-time, the respiratory time. And so the respiratory rate goes up and decreases the auto-keeper. So I will say this. Honestly, eye time doesn't really matter too much.
and decreasing it may lead to unpredictable responses. Sometimes they may be good, but sometimes it actually works in the patients. So I'm almost at the end. What initial evaluator settings are best? Again, aim to meet your goals of ventilation and oxygenation. This is my personal approach, and I get this from literature from Toby.
So the cyst control tidal volumes don't have to be close to the 60 cc per kg. I like to use somewhere around 8 generally, and a lot of times I'm able to go much higher than that because of the lung compliance being so good. Use respiratory rates a little bit on the lower side. I like decelerating way forward over square and spaces like that anyways. And again, flow rates don't have to be exceedingly high to decrease your respiratory time. 60 as a baseline should be okay.
I always use a PEEP and BF-Io2 usually isn't going to be that high in these patients. So we do as matching as if although the present doesn't cause necessarily cosyloxemia. And then for an acid-base... goals and targets always aim to get acceptable numbers and not normal numbers, try to get the pH to 7.2 or above 7.2. In our patient, when they first got intubated, they were profoundly acidic and they could notice the tidal volume was a little bit on the lower side despite, actually, also probably worsened by the decreased respiratory rate that they issued per month. When the ventilator incentives adjusted less than 700, they'll notice the acid status of the patient.
In Hesse's textbook, Mechanical Violation, there's nice algorithms that you can potentially look at to basically explain everything that we've been talking about if you aim to meet your goals, adjustments, that's a good mean in general. I'm not going to go through all of these, of course. You can take a look at them if you're able to actually get my talk as well. But the most important thing is balance, right? So you want to be somewhere in the middle. You want to give a...
Try to achieve a low-minute ventilation. You don't need to achieve normal PCO2s. Try to get moderate sedation. And people sometimes need to be paralyzed. And if you can paralyze less, the better. I will say that asthma patients probably get paralyzed a lot more than COPD patients and they require a lot more sedation in general. Asthma patients fall into two categories. There's your acute bronchospasm that goes away quickly. But you also have your long-prolonged airway inflammation, which can be a little bit more difficult to treat. And these are the patients that have a sedation paralysis.
Avoid the hypotension barotrauma because that's what suits high mortality, but keep also in mind that if you paralyzed patients too much, you will cause myopathy as well. These are my take-home points. Use the load that you'd like. Again, acceptable creation, not normal, because CO2, respiratory rate going down is best, and then external keep as needed, but don't overdo it. It improves work of breathing, but not the dynamic population. I'm sorry about that. I actually went nine minutes above.
I hope this was helpful for everybody. I'm happy to take any questions. I will stare at the chat box as well. Again, this is the QR code, my email, my number. I'm happy to connect with anybody if you guys have any questions or comments.